20KW High Frequency X-ray Machine JY-A27
1.Maximum output power
?Intermittent mode: 100kV ,200mA,20kW。
2.Adjustment range of tube current、tube voltage
?tube current: Radiography 16mA~200mA。
?tube voltage: Radiography 40kV~125 kV。
3.Exposure Time:photography0.0025s~6.4s step adjustable
4 Nominal Frequency≥30kHz
5.The condition of power
?Power Voltage:380V±10%
Power supply capacity: 120% the power of the equipment
?Internal resistance:≤0.5Ω。
6.High frequency and high voltage x-ray generator(including microcomputer control console)
?Nominal frequency of high voltage x-ray generator≥30kHz。
?Internal resistance: less than 0.5Ω.
?Nominal Output power:20kW。
?Output Voltage: Radiography: 40 kV~125kV, step 1kV;
?Output Current: Radiography: 16mA~200mA,
?Exposure Time: 0.0025s~6.4s。
?Display:digital pipe display photography voltage\ Photography current\ photography time\ mAS of photography
?X-ray tube
?Max tube voltage:125kV Max tube current:400mA。
?Small focus:1.0mm ;Big focus:2.0mm。
?Speed of anode rotating:2800r/min
?Anode heat capacity:140KHU
7.Collimator
?Collimator with fixed position lamp.
?Power supply: Working voltage: 24V
8.Radiography table
?Integration of the table with stand column without skyrail design.
?Length*width*height of table surface:2000mm×760mm×690mm. Deviation:±10mm
?Distance from table surface to film:50mm
?Horizontal movement range of table≥220mm
?Vertical movement range of table≥900mm
has locking function, electromagnetic braking, Braking force: ≥98N
9.X-ray source component pillar:
?With scale that can indicate the height of the x-ray tube unit
?Distance of vertical movement:770mm±10mm;+90°~-90°, deviation<±2°
?Moving crosswise (left & right):+180°~-180°, deviation<±2°
?Moving lengthways (along ground railway) ≥1300mm
10.Bucky: Fixed beam filtration grid density≥40 lines/cm.
grid ratio 10:1,。Focus Distance:110cm
11. Horizontal movement of bucky ≥500mm
Electromagnetic braking
Braking force ≥98N,
Button control
12. Collimator open and closedown should be flexible, reliable,can adjust to any Radiation in field rectangle size.
13. Stand chest stand
Fixed beam filtration grid density≥40 lines/cm.
Ray-filter grid:grid ratio 10:1, Focus Distance:110cm
Cassette size:5″×7″ 8″×10″ 10″×12″ 11″×14″ 14″×17″